Beamex EXTB

EXTERNAL PRESSURE MODULE

DAILY RATES
Cost per Day $100.00 $85.00 $75.00 $60.00
Duration (days) 1-7 8-14 15-29 30
TYPICAL HIRE PERIODS
Cost $700.00 $1,190.00 $2,175.00 $1,800.00
Duration 7 Days 14 Days 29 Days 30 Days
Committing to 3, 6 or 12 months?
Contact us on 1300 623 605 for fantastic long term hire rates.

Temperature coefficient±0.001% Rdg/ °C outside 15…35 °C (59…95 °F)

INT10mD / EXT10mD < ±0.002% Span/ °C outside 15…35 °C (59…95°F)

  1. IS version available for all external pressure modules
  2. Every internal/external pressure module’s range may also be displayed in absolute pressure if the Barometric
    Module (B) is installed.
  3. ‘Accuracy’ includes hysteresis, nonlinearity, repeatability and reference standard uncertainty (k=2).
  4. ‘1 Year Uncertainty’ includes hysteresis, nonlinearity, repeatability and typical long-term stability for
    mentioned period (k=2).

All external pressure modules (EXT) are also compatible with Beamex MC6, MC5, MC4, MC2 and MC5P calibrators. All
intrinsically safe external pressure modules
(EXT-IS) are compatible with Beamex MC5-IS and MC2-IS calibrators.

Supports the following pressure units as standard:

Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at, mmH2O, cmH2O, mH2O, iwc, ftH2O,
mmHg,
cmHg, mHg, inHg, mmHg(0 °C), inHg(0 °C),
mmH2O(4 °C), inH2O(4 °C), ftH2O(4 °C), inH2O(60°F), mmH2O(68°F), inH2O(68°F), ftH2O(68°F), torr, atm.
INT B / EXT B; M5 (10/32”) female.
INT10mD and EXT10mD; Two M5 (10/32”) female threads with a hose nipple included.
INT100m/EXT100m – INT20C/EXT20C; G1/8” (ISO228/1) female. A conical 1/8” BSP male with 60° internal cone adapter
included for Beamex hose set.
INT60, INT100, INT160; G1/8” (ISO228/1) female. EXT60, EXT100, EXT160, EXT250, EXT600, EXT1000; G ¼” (ISO228/1)
male.
Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.
Maximum overpressure;
B module; 1200 mbar abs. 10mD module; 200 mbar. EXT600; 900 bar. EXT1000; 1000 bar.
For all other modules, the maximum overpressure is twice the nominal range.
HART is a registered trademark of HART Communication Foundation.

Related Products

Success message!
Warning message!